We use cookies in order to guarantee the best possible service. If you continue browsing the site you consent to our cookie use.

CableEye tester for cables and cable harnishes

CAMI/CableEye Test System for Cables and Harnesses

Meilhaus Electronic - Competence in CableEye Cable Testers - Since 2010!

The Expandable Cable Tester!

 


  M2U-B M2U M3U M3UH M4 HVX HVX-21
810U 811U 821U 821UH 824 829 829A
Continuity
Opens
Short
Miswires
Intermittent faults
Resistance
Connection/non-connection quality
Resistance (2-wire)
Continuouse resistance scan
Fixture resistance nulling
Resistance (4-wire Kelvin) () optional () optional
Capacitance
Wire length, cable length () optional () optional
Length to break () optional () optional
Twisted-pairing () optional () optional
Single channel safety test
Chassis, panels, transformers etc.
Insulation quality
Dielectric strenght,
DWV/Dielectric Withstand Voltage
Insulation resistance
Inline components
Resistors
Diodes          
   Orientation
   Forward voltage
LEDs          
   Orientation
   Colour detection
Zener diodes          
   Orientation
   Forward voltage
   Reverse voltage <10 V
   Reverse voltage >10 V
Capacitors () optional () optional
  M2U-B M2U M3U M3UH M4 HVX HVX-21
810U 811U 821U 821UH 824 829 829A
Test points1) 128 152 152 152 152 152 for LV-Test,
128 for HV-Test
152 for LV-Test,
128 for HV-Test
Expandable - ✓, to max. 1024 ✓, to max. 2560 ✓, to max. 1024 ✓, to max. 1024 ✓, to max. 1024 ✓, to max. 512
Test time2) 0.2 s/- 0.2 s/- 0.2 s/0.25 s 0.2 s/0.25 s 0.15 s/0.4 s depending on voltage, ramp rate, test algorithm selected depending on voltage, ramp rate, test algorithm selected
Resistance threshold 46 kΩ fixed 46 kΩ fixed 0.3 Ω...10 MΩ 0.1 Ω...5 MΩ 0.02 Ω...10 MΩ 0.1 Ω...1 GΩ 0.1 Ω...5 GΩ
Resistance accuracy - - 2% (10 Ω...100 Ω);
1% (100 Ω...1 MΩ);
lesser accuracy over full range
2% (10 Ω...100 Ω);
1% (100 Ω...1 MΩ);
lesser accuracy over full range
2% (10 Ω...100 Ω);
1% (100 Ω...1 MΩ);
lesser accuracy over full range
same as M3UH under 1 MΩ;
5% (1 MΩ...100 MΩ);
lesser accuracy over 100 MΩ
same as M3UH under 1 MΩ;
5% (1 MΩ...100 MΩ);
lesser accuracy over 100 MΩ
4-wire Kelvin - - - - 20 mΩ ±20 mΩ,
20 mΩ...15 Ω,
test current 3.3 mA
option: 1 mΩ ±1 mΩ,
1 mΩ...15 Ω,
test current 100 mA...1 mA
option: 1 mΩ ±1 mΩ,
1 mΩ...15 Ω,
test current 100 mA...1 mA
Intermittent connection scan rate 33 scans/s 33 scans/s 26 scans/s (128 test points), 90 scans/s (64 test points) 26 scans/s (128 test points), 90 scans/s (64 test points) 18 scans/s (128 test points), 47 scans/s (64 test points) same as M3UH or M4 same as M3UH or M4
Diode measurement Orientation only Orientation only Orientation, forward voltage, reverse breakdown <10 V Orientation, forward voltage, reverse breakdown <10 V Orientation, forward voltage, reverse breakdown <10 V Orientation, forward voltage, reverse breakdown >10 V Orientation, forward voltage, reverse breakdown >10 V
Test voltage 5 V 5 V 10 V 10 V Settable 10 V (default), 5 V, 3.3 V, 2.5 V, 1.7 V 10...1500 VDC or 10...1000 VACRMS in increents of 1 V 10...2100 VDC or 10...1200 VACRMS in increments of 1 V
Capacitance - - - - Range 100 pF...100 µF, 5% accuracy, measurement rate 20 measurements/s at 100 nF or less Range 100 pF...100 µF, 5% accuracy, measurement rate 20 measurements/s at 100 nF or less Range 100 pF...100 µF, 5% accuracy, measurement rate 20 measurements/s at 100 nF or less
Twisted pair measurement - - - - ✓, approx. 183 cm min. length ✓, approx. 183 cm min. length ✓, approx. 183 cm min. length
Measurement cable length - - - - ✓, approx. ±91 cm ✓, approx. ±91 cm ✓, approx. ±91 cm
Measurement distance to break - - - - ✓, min. distance to break approx. ±91 cm ✓, min. distance to break approx. ±91 cm ✓, min. distance to break approx. ±91 cm
Dwell time range 1 µs...100 ms 1 µs...100 ms 1 µs...100 ms 1 µs...100 ms 1 µs...100 ms LV: 1 µs...100 ms,
HV: 30 ms...300 s
LV: 1 µs...100 ms,
HV: 30 ms...300 s
Insulation resistance measurement - - 10 MΩ at 10 V 5 MΩ at 10 V 10 MΩ at 10 V 2 MΩ...1 GΩ (1500 VDC);
2 MΩ (min., 1000 VAC);
current sensitivity 1 µA
2 MΩ...5 GΩ (2100 VDC);
2 MΩ (min., 1000 VAC);
current sensitivity 0.2 µA
Digital I/Os Inputs only Pairs of test points used as inputs, 50+ relay outputs Pairs of test points used as inputs, 50+ relay outputs Pairs of test points used as inputs, 50+ relay outputs Pairs of test points used as inputs, 50+ relay outputs Pairs of test points used as inputs, 50+ relay outputs Pairs of test points used as inputs, 50+ relay outputs
Calibration - - Recommended yearly Recommended yearly Recommended yearly Recommended yearly Recommended yearly
Test point connectors 64-pin dual-row headers, 2.54 mm centers; 2 per 128-point module 64-pin dual-row headers, 2.54 mm centers; 2 per 128-point module 64-pin dual-row headers, 2.54 mm centers; 2 per 128-point module 64-pin dual-row headers, 2.54 mm centers; 2 per 128-point module 64-pin dual-row headers, 2.54 mm centers; 2 per 128-point module 64-pin dual-row headers, 2.54 mm centers; 2 per 128-point module 64-pin dual-row headers, 2.54 mm centers; 2 per 128-point module
Remote control socket - MiniDIN8 connector for footswitch, ext. control panel MiniDIN8 connector for footswitch, ext. control panel MiniDIN8 connector for footswitch, ext. control panel MiniDIN8 connector for footswitch, ext. control panel MiniDIN8 connector for footswitch, ext. control panel MiniDIN8 connector for footswitch, ext. control panel
Probe socket - ✓, probe included with tester ✓, probe included with tester ✓, probe included with tester ✓, probe included with tester ✓, probe included with tester ✓, probe included with tester
Power requirements 9 VDC at max. 300 mA, 3 W (ext. power supply) 9 VDC at max. 300 mA, 3 W (ext. power supply) 18 VDC at max. 500 mA, 9 W (ext. power supply) 18 VDC at max. 500 mA, 9 W (ext. power supply) 18 VDC at max. 500 mA, 9 W 100...250 VAC/50...60 Hz, 130 W (max.) for 128 test points; 175 W (max.) for 512 test points; power cord 100...250 VAC/50...60 Hz, 130 W (max.) for 128 test points; 175 W (max.) for 512 Test points; power cord
Weight 1.1 kg 1.1 kg 1.2 kg 1.2 kg 1.1 kg 9.5 kg 9.5 kg
PC requirements Windows XP SP3 or later; compatible with touchscreen and laptop PCs Windows XP SP3 or later; compatible with touchscreen and laptop PCs Windows XP SP3 or later; compatible with touchscreen and laptop PCs Windows XP SP3 or later; compatible with touchscreen and laptop PCs Windows XP SP3 or later; compatible with touchscreen and laptop PCs Windows XP SP3 or later; compatible with touchscreen and laptop PCs Windows XP SP3 or later; compatible with touchscreen and laptop PCs
USB interface USB 1.1 full-speed USB 1.1 full-speed USB 1.1 full-speed USB 1.1 full-speed USB 1.1 full-speed USB 1.1 full-speed, 2 Ports USB 1.1 full-speed, 2 Ports

1) LV = low voltage test, HV = high voltage test.
2) 128 test points, first value continuity only, second value with resistance test.

Compatibility Between CB Boards and CableEye Testers

A green checkmark denotes that the CB board is compatible with the tester displayed in the column.
A red x denotes that the CB board is NOT compatible with the tester displayed in the column.
A number in the "Expansions Required" column means that the tester requires the number of expansion modules displayed to use the board.
Only for the HVX High Voltage testers, the "Voltage Rating" column displays the maximum voltage that the board can sustain without dielectric breakdown.

CB Board M2U Basic M2U / M3U / M4 / HVX Expansions Required Voltage Rating
CB1 500
CB2 500
CB2A 500
CB3 500
CB4 500
CB5 500
CB6 500
CB7 1 500
CB8 500
CB9 500
CB10 1 500
CB11 500
CB12 60 Conductors Max 60 Conductors Max 1 500
CB12A 60 Conductors Max 60 Conductors Max 1 500
CB13 500
CB14 500
CB15 500
CB16 500
CB16A 500
CB17 500
CB18 500
CB18A 500
CB18B 500
CB18C 500
CB19 500
CB20 1 500
CB21 1 500
CB22 500
CB23 500
CB24 1 500
CB25 500
CB25A 500
CB25B 500
CB26A 500
CB26B 500
CB26C 500
CB26D 500
CB26E 500
CB26F 500
CB26G 500
CB26H 500
CB26I 500
CB26K 500
CB26L 500
CB26M 500
CB26N 500
CB26T 500
CB27 500
CB28 500
CB29 500
CB29A 500
CB29H 500
CB29HA 500
CB29BB 500
CB30 1 500
CB30A 1 500
CB30B 2 500
CB30C 2 500
CB31 1 500
CB32 1 500
CB33 500
CB34 500
CB35 500
CB37A M3U/M4/HVX 500
CB38A M3U/M4/HVX 1 500
CB39 500
CB40 500
CB41 500
CB42 500
CB43 500
CB44 500
CB45 500
CB46 1 500
CB47 500
CB48A 500
CB48B 500
CB49 500
CB50 M3U/M4/HVX 500
CB-T1 500

 


Basic LV and HVX Systems

 


Expansions

 


Connector Boards and Accessories (Excerpt)

 

Couldn't find the desired product in the shop? Just ask us - more products available!