Benefits of the LR8450(-01) Automotive and General Purpose Dataloggers

  • Main device with display and up to 4 plug-in and/or 7 wireless I/O modules.
  • Covers all important I/O signals for industrial and automotive applications incl. CAN.
  • Flexible distributed I/O with wireless modules (LR8450-01).

HIOKI LR8450 Highly Flexible Modular Dataloggers up to 1 kS/s

The HIOKI LR8450(-01) series are portable modular dataloggers at max. 1 kS/s (1 ms) for voltage, strain, temperature, humidity, resistance and CAN. The loggers consist of the main device with user interface (buttons, display etc.) and the I/O modules. The LR8450 is the standard model designed for use with plug-in modules only. The LR8450-01 is the wireless version for additional WLAN modules. This modular design allows very flexible and cost effective logging applications in automotive, industrial and other fields, also with distributed I/O through the wireless I/O modules (LR8450-01).

  • LR8450:
    • Standard model, designed for use with plug-in modules only.
    • 6 module types available for voltage, temperature, rel. humidity, strain, resistance, and CAN.
  • LR8450-01:
    • Wireless LAN model.
    • Add up to 7 wireless modules in total for a maximum of 330 channels.
    • Mix plug-in and wireless modules: Add channels freely via either plug-in or wireless modules, or use exclusively with wireless modules.
    • 6 module types available for voltage, temperature, rel. humidity, strain, resistance, and CAN.
  • Convenient operation of the main unit with large display, storage on SD cards/USB stick, LAN and USB interface and many useful functions such as triggers, alarms and calculations/signal operations.

Model Overview Dataloggers LR8450 and LR8450-01 Wireless

ModelLR8450LR8450-01
TypeData loggers/memory loggers main device - the LR8450 and LR8450-01 cannot perform measurement on their own. One or more plug-in modules or wireless modules are required (sold separately)
Max. number of modules4 plug-in modules (max. 4 CAN modules U8555)4 plug-in modules and/or 7 wireless modules (max. 4 CAN modules U8555 and/or LR8535)
ChannelsMax. 120 channelsMax. 330 channels (120 channels + wireless 210 channels)
MeasurementsDepending on module(s): DC voltage, temperature/thermocouple sensor, humidity, cumulative pulses, rotation, digital, resistance measurement, strain
RecordingMin. interval 1 ms, max. interval 1 h; realtime recording to SD card (2 GB/8 GB)/USB memory stick; internal buffer memory volatile, 256 Mwords
Additional functionsCalculations (numeric, waveform calculations), triggers, alarms and more
Display7"/17.8 cm TFT color LCD (WVGA 800x480 dots) with backlight
InterfacesUSB 2.0 host and device, LAN (100Base-TX/1000Base-T)USB 2.0 host and device, LAN (100Base-TX/1000Base-T), Wireless LAN (IEEE 802.11b/g/n; communications range 30 m, line of sight)
Power supplyAC adaptor Z1014; battery: 2 batteries Z1007 battery pack (when used with AC adaptor, AC adaptor takes priority) Li-ion, 7.2 V, 2170 mAh; external power supply 10...30 VDC
Dimensions (mm)Without modules: 272 x 145 x 43; with 2 modules: 272 x 198 x 63; with 4 modules: 272 x 252 x 63; aoprox. 1108 g (main device only, excluding battery pack)

Model Overview U8550, U8551, U8552, U8553, U8554, U8555 Datalogger Plug-in Modules

Plug-in datalogger unitsFor logger
U8550Voltage/temperature unit: 2 terminals M3 mm screw type, 15 channels, voltage, temperature with thermocouple, or humidity measurement, max. 10 ms/S; 16 bit A/DLR8450 and LR8450-01
U8551Universal unit: 4 terminals push-button type, 15 channels, voltage, temperature with thermocouple, Platinum resistance temperature sensor, humidity, or resistance measurement,max. 10 ms/S; 16 bit A/DLR8450 and LR8450-01
U8552Voltage/temperature unit: 2 terminals push-button type, 30 channels, voltage, temperature with thermocouple, or humidity measurement, max. 20 ms/S (10 ms at 15 channels or less); 16 bit A/DLR8450 and LR8450-01
U8553High speed voltage unit: 2 terminals M3 mm screw type, 5 channels, voltage, max. 1 ms/S; 16 bit A/DLR8450 and LR8450-01
U8554Strain unit: Strain gauge converter input, 5ch, bridge box included, max. 1 ms/S; 16 bit A/DLR8450 and LR8450-01
U8555CAN unit: CAN/CAN FD input and output switchable, 2 channels, max. 10 ms/S (up to 50 CAN signals), max. 500 CAN signals (at 100 ms/S)LR8450 and LR8450-01
U8556Current measurement module, 5 channels, max. sampling rate 1 ms, 16 bit resolution; HIOKI PL14 connector; ideal for use with current sensors CT7812/CT7822 und others (supported sensors see datasheet)LR8450 and
LR8459-01
U855715 channels, temperature (thermistor) and resistance measurement: Convert and display the temperature value for any thermistor up to 200 kΩ; sampling rate of up to 50 ms with 15 channels; resistance measurement (resolution): 2000 Ω (0.1 Ω), 20 kΩ (1 Ω), 200 kΩ (10 Ω); temperature conversion methods: Steinhart–Hart equation, B-constant (Beta parameter) equation; resolution: 0.01°C; temperature range: -100...+300.0°CLR8450 and
LR8459-01

Model Overview LR8530, LR8531, LR8532, LR8533, LR8534, LR8535 Datalogger Wireless Modules

Wireless datalogger unitsFor logger
LR8530Wireless voltage/temperature unit: 2 terminals M3 mm screw type, 15 channels, voltage, temperature with thermocouple, or humidity measurement, max. 10 ms/SLR8450-01
LR8531Wireless universal unit: 4 terminals push-button type, 15 channels, voltage, temperature with thermocouple, Platinum resistance temperature sensor, humidity, or resistance measurement, max. 10 ms/SLR8450-01
LR8532Wireless voltage/temperature unit: 2 terminals push-button type, 30 channels, voltage, temperature with thermocouple, or humidity measurement, max. 20 ms/S (10 ms at 15 channels or less)LR8450-01
LR8533Wireless high speed voltage unit: 2 terminals M3 mm screw type, 5 channels, voltage, max. 1 ms/SLR8450-01
LR8534Wireless strain unit: Strain gauge converter input, 5 channels, bridge box included, max. 1 ms/SLR8450-01
LR8535Wireless CAN unit: CAN/CAN FD input, 2 channels, max. 10 ms/S (up to 50 CAN signals), max. 500 CAN signals (at 100 ms/S)LR8450-01
LR8536Wireless measurement module, 5 channels, max. sampling rate 1 ms, 16 bit resolution; HIOKI PL14 connector; ideal for use with current sensors CT7812/CT7822 und others (supported sensors see datasheet)LR8450-01
LR853715 channels, temperature (thermistor) and resistance measurement: Convert and display the temperature value for any thermistor up to 200 kΩ; sampling rate of up to 50 ms with 15 channels; resistance measurement (resolution): 2000 Ω (0.1 Ω), 20 kΩ (1 Ω), 200 kΩ (10 Ω); temperature conversion methods: Steinhart–Hart equation, B-constant (Beta parameter) equation; resolution: 0.01°C; temperature range: -100...+300.0°CLR8450-01

For detailed specifications of the loggers and I/O modules check the full datasheet.

Included: Logger LR 8450 or 8450-01, quick start manual, LOGGER application CD (quick start manual, instruction manual, logger utility, logger utility instruction manual, CAN editor, CAN editor instruction manual, communication instruction manual), USB cable, AC adaptor Z1014, precautions concerning use of equipment that emits radio waves (LR8450-01 only).

Frequently Asked Questions:

Question: For the HIOKI data logger LR8450, there are modules available for different temperature sensors. When to use which model?

Answer: Sensor types: The choice primarily depends on the sensors you are using. Thermocouples (various types such as J, K, T, E, R, S, B, N) are commonly used in industrial processes, combustion systems, aerospace testing, etc. They offer wide temperature ranges up to thousands of °C, are fast, and cost-effective. However, compared to other sensor types, their accuracy is lower.
RTDs (types such as Pt100, Pt1000) are more expensive and slower, but very accurate and offer high long-term stability. Therefore, they are mostly used in climate chambers, calibration laboratories, process control, etc.
Thermistors (NTC, PTC – negative or positive temperature coefficient) are fast and inexpensive, but have a comparatively limited temperature range. They are commonly used in consumer electronics/household applications, EV batteries, car engines, air conditioning systems, etc.

Spatial arrangement: The HIOKI I/O modules of the U series are attached directly to the base data logger, while the modules of the LR series are connected wirelessly to the base data logger. For widely distributed measurement setups, the LR modules offer several advantages: placing the I/O modules close to the measurement points ensures short sensor leads, resulting in less susceptibility to interference, reduced noise, and improved accuracy. Distributing I/O modules across different measurement locations is very convenient. In battery operation, complete autonomy can be achieved.

Additional measurements (if required): Depending on the variant, the I/O modules also support other measurement parameters: voltage (0...100 V), resistance, humidity, 4…20 mA, and heat flow.

Comparison of Temperature Modules

ModelU8550 | LR8530
voltage/temperature modules
U8551 | LR8531
universal modules
U8552 | LR8532
voltage/temperature modules
U8557 | LR8537
thermistor modules
Voltage measurement0...100 V, 10 ranges0...100 V, 10 ranges0...100 V, 10 ranges-
Temperature measurementThermocouplesThermocouples, RTDsThermocouplesThermistors up to 200 kΩ
Resistance measurement-0...200 Ω, 4 ranges-0...200 kΩ, 3 ranges
OtherHumidity, 4...20 mA, heat flowHumidity, 4...20 mA, heat flowHumidity, 4...20 mA, heat flow-
Channels15153015
Update interval10 ms10 ms10 ms/20 ms50 ms

Information on product safety:

Manufacturer:
HIOKI E.E. Corporation, 81 Koizumi Ueda, Nagano 386-1192/JPN
www.hioki.com
(EU branch Germany)